Advanced Piezoelectric Testing

 
 
The Piezo Task adds to the standard Hysteresis Task automatic measurement of the output of a displacement meter to detect the motion at the surface of a Piezoelectric sample as it is stimulated by tester. The displacement meter output is connected to the SENSOR input at the rear of the Precision tester . Radiant has combined the original PiezoTask with specialized math filters to create this more advanced acquisition task that will accurately measure extremely small displacement butterfly loops. When the displacements being measured are so small, less than 300 nanometers, ambient environmental effects from vibration to air currents to thermal expansion of test fixtures will distort the results. Averaging alone will not do. Averaging alone simply reduces the effects of the environmental distortions on the final measurement. Instead, each loop must be zero'd and de-trended individually before averaging and smoothing. The Advanced Piezo Task performs these actions as it collects the displacement data. Since the Precision Testers will accept displacement data from any displacement sensor, including interferometers and AFMs, Advanced Piezo makes it possible to directly measure the piston motion of thin ferroelectric films. Bulk Ceramics & MEM's can also be measured with the same software.
 
For detailed information, read the technical instructions for connecting the displacement sensor to the SENSOR input of the Radiant tester, Using the Sensor Port, and then learn how to collect and process displacement data with the Application Note for Advanced Piezo.
 

Measurements Include

 
 
The following measurements can be taken with Piezo Task:
  • Large Signal Polarization and Displacement
  • Large Signal vs. Small Signal Measurement
  • Piezoelectric vs. Unipolar, Bipolar, and DC Bias Voltages
  • Direct D33 and D31 Measurements
  • Capacitance and Piezoelectricity
  • Hysteresis and Displacement
 

Displacement Sensors

 
Any displacement sensor that outputs its measurements as a voltage may be connected to a Precision tester to capture the movements of a ample during a measurement.  Radiant has extensive experience with the following sensor units:
 
 
Parameter:

 

Instrument

 

Scale

Factor

Absolute Resolution on

Premier II

Noise Limited Resolution Premier II

Absolute Resolution on

LC

Noise Limited Resolution LC

DI Dimension 3100 AFM

100Å/V

0.008Å

0.10Å

0.031Å

0.12Å

Seiko Epson SPA 400 AFM

50Å/V

0.004Å

0.05Å

0.015Å

0.06Å

Asylum SA AFM

700Å/V

0.053Å

7.00Å

0.214Å

0.84Å

MTI2100 2032RX Photonic Sensor

5000Å/V

0.380Å

5.00Å

1.525Å

6.00Å

MTI2100 2032R    Photonic Sensor

19m/V

14.44Å

190Å

57.95Å

228Å

Polytec OFV534 Laser Vibrometer

500Å/V

0.038Å

0.50Å

0.153Å

0.60Å

SIOS SP-S 120 Laser Vibrometer

2400Å/V

0.182Å

2.40Å

0.732Å

2.88Å