Precision LC II Ferroelectric Tester

An ideal, affordable, general purpose tester.

The Precision LCII is Radiant's answer to a growing need for affordable characterization equipment. It is an ideal general-purpose tester with a broad test range for thin films and bulk ceramics, with a frequency rating of 5KHz at ±200V built-in to the system.

5KHz @ ±200V

Maximum hysteresis frequency

<10fC

Minimum charge resolution

18

Number of ADC bits

10kV

With an external amplifier and HVI

Precision LC II ferroelectric tester with a High Voltage Interface on top

Precision LC II (bottom) with a High Voltage Interface (top).

Resolution and range

Using up to 32,000 points with 18-bit resolution from 2kHz out to 1/30th of a Hertz for capacitors ranging from square microns to square centimeters, the LC II enables high resolution characterization for bulk capacitors.

Built-in drive options

The Precision LCII is offered with a variety of internal amplifiers. The Precision LCII is offered in a ±10V, 30V, 100V, 200V, and 500V built-in drive volt option. The Precision LCII can be expanded to 10kV with the addition of a high voltage interface and amplifier.

Measurements without reconnecting

The Precision LCII executes hysteresis, pulse, leakage, IV, and CV (and many other tasks) measurements without changing sample connections. With the addition of extra fixtures, the Precision LCII can measure pyroelectric properties, magnetoelectric properties, transistor characteristics, cryogenic properties, bulk and/or thin film piezoelectric properties.

Specifications

Precision LC II

Drive voltage

Voltage Range(built-in drive voltage) ±10V, ±30V, ±100V, ±200V, or ±500V built-in
Voltage Range with an external amplifier and high voltage interface (HVI) 10KV

Charge and area resolution

Number of ADC Bits 18
Minimum Charge Resolution <10fC
Minimum Area Resolution (assuming 1 ADC bit = 1μC/cm²) <1μ²
Maximum Charge Resolution 276μC
Maximum Area Resolution (assuming saturation polarization = 100μC/cm²) 2.76cm²
Maximum Charge Resolution with High Voltage Interface (HVI) 27.6mC
Maximum Area Resolution (assuming saturation polarization = 100μC/cm²) w/ HVI >100cm²

** Maximum hysteresis frequency applies to the internal amplifier purchased: 5KHz with the ±10V, ±30V, ±100V, or ±200V amplifier, and 2KHz with the ±500V amplifier.

Hysteresis and pulse timing

Maximum Hysteresis Frequency** 5KHz @ 10V
5KHz @ 30V
5KHz @ 100V
5KHz @ 200V
2KHz @ 500V
Minimum Hysteresis Frequency 0.03Hz
Minimum Pulse Width 50μs
Minimum Pulse Rise Time (5V) 40μs
Maximum Pulse Width 1s
Maximum Delay between Pulses 40ks
Internal Clock 25ns
Output Rise Time Control 103 scaling

Leakage, small signal, and input

Minimum Leakage Current (assuming max current integration period = 1 seconds) 12pA
Maximum Small Signal Cap Frequency 20KHz
Minimum Small Signal Cap Frequency 1Hz
Input Capacitance ~6fF
Electrometer Input All Test Frequencies for all test at any speed Yes

Price an LC II for your lab.

Vision Software is included with every Precision test system.

Radiant’s Precision Line of Test Systems are driven by Vision Software, with a library of over 150 measurement, analysis, plotting, program control, and documentation tasks.

Combine tests in one run

Chain Hysteresis, PUND, Fatigue, and Leakage in one Test Definition instead of running one test at a time.

Control the environment

Set temperature over GPIB, set the field of a Quantum Design superconducting magnet, and query an LCR meter within the test.

Let the logic decide

Branch tasks repeat steps until a value reaches your threshold; If/Then and Exit tasks skip steps or end the run based on a measured value.

Nothing gets lost

Every run is archived with its Test Definition, so any experiment can be recalled or reproduced.

Documentation and support

The LC II brochure and the company brochure, plus the full knowledge base library.

RT shared styles: global CSS for the redesign. Edit this via Edit Template. This label is hidden on the live site.